Título : Virtual goniometer using 3 space mocap sensors for lower limbs evaluation
Autor : Buele, Jorge
Pilatásig, Marco
Angueta, Hamilton
Ruales-Martínez, María-Belén
Varela-Aldás, José
Fecha de publicación : 2020
Editorial : Advances in Intelligent Systems and Computing. Volume 1137 AISC, Pages 439 - 448. International Conference on Information Technology and Systems, ICITS 2020. Bogota. 5 February 2020 through 7 February 2020
Resumen : The deterioration of motor skills in lower limbs represents a reduction in people’s quality of life. This paper presents a virtual goniometer developed by the 3 Space Mocap sensors and Unity 3D software. The system shows lower limbs angles in real-time. This system helps physiatrists determine the movements of hip and knee and was compared with a goniometer to corroborate the system performance. To validate this prototype, comparative tests are carried out present a standard deviation of 1.86° and 2.12° in the left hip and knee respectively. Therefore, it confirms to be a reliable proposal for physical evaluations of lower limbs. © Springer Nature Switzerland AG 2020.
URI : https://link.springer.com/chapter/10.1007/978-3-030-40690-5_43
http://repositorio.uti.edu.ec//handle/123456789/3396
Aparece en las colecciones: Artículos Científicos Indexados